18
“CS Unit”
– PSI or MPa
“Temp Unit”
– °F or °C
“Archive Path”
– Choose a location to save archive files.
“Logo Path”
– Select a logo (.JPG format) to print on the graph at the
end of the test.
“CS #1”, “CS #2”, “CS #3”
– Enter a compressive strength value in each
of these fields. When the cement reaches that compressive strength, the
software will record the elapsed time and print it on the graph.
“Unit #” –
Identifies the unit generating the graph. This field is helpful
when multiple units are generating graphs.
“Strength @”, “Strength 2 @”, “Strength 3 @”
– Enter a time period in
each of these three fields. At the end of the test, the compressive strength
at these three times will print on the graph.
“Ultrasonic Transducer”, “Pressure Transducer”, “Thermocouple”
–
Calibration Frequency:
Enter a time period in each of these three fields.
The default time period is set at six months but can be edited if needed.
The UCA software will prompt a reminder that these components need to
be recalibrated at each of these times. This will not affect the test.
“Print to Printer”
– When this option is on, a graph of the test results
will automatically print to the computer's default printer when a test is
complete.