Top Cell Cap
Top Cap Assembly
1.
Inspect the transducer, fill, and thermocouple ports and make sure they
are clean and free of debris. Dried transducer couplant can accumulate
on the sides and bottom of the port. This couplant must be cleaned from
the port as well.
The two ports are interchangeable.
2.
Tighten the jaws of the vice on the flats of the top cell cap with the threads
facing up.
3. Apply a thin layer of high-temperature grease to the threads of the bottom
cell cap.
4. Remove grease from the center area of the surface with a paper towel.
5. Leave only a thin layer of grease on the top surfaces of the bottom cell
cap avoiding the center area that transmits or receives transducer signal.
Excess grease will interfere with the transducer signal.
6. Apply high temperature ultrasonic couplant (#123-024) to the internal
center of the cell cap where the excess high temperature grease was
removed.
High Temp Grease
Application Area
Ultrasonic Couplant
Application Area