OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
35
SGSM
Software Setup
Before you begin an SGSM test, you must prepare the PC to record the data.
1. Turn on the PC and UCA.
2. Put the UCA/SGSM switch to the SGSM position.
3. Open the software by double-clicking the automated SGSM icon on the
desktop.
4. Click Utilities
→
Setup.
The SGSM software will only communicate through the serial connection.
COM9
C:\Automated SGSM Data
C:\Program Files (x86)\SGSM\logos\OFITE Logo (OFI Testing, Equipment, Inc. Tag) -
lb/100ft^2
6
6
“COM Port”
- The COM port is connected to the device through the
RS-232 cable.
“Temp Unit”
- °F or °C
“Pressure”
- PSI or MPa
“Gel Strength”
- Gel strength units (Dyne/cm
2
, lb/100 ft
2
, Pa., lb/ft
2
)