OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
27
10. Remove the top cell cap from the vise and replace it with the bottom cell
cap and cell body. Tighten the jaws of the vice on the flats of the cell cap
and cell body.
7. Install the seal ring (#120-50-027-1) with the narrow side pointed towards
the threads and place the o-ring (#123-011) on top of it.
The o-ring should be inspected for signs of degradation
before installing a new o-ring.
8. Install the retaining ring (#120-50-026) above the o-ring.
9. Apply more high temperature grease to the seal ring,
o-ring, and the retaining ring.
Top Cell Cap
(#120-50-021C)
Retaining Ring
(#120-50-026)
O-ring
(#123-011)
Seal Ring
(#120-50-027-1)
Seal Ring
(Cross Section)