OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
61
19. Remove the cement from the bottom of the plug in small chunks until you
expose one of the vanes on the bob.
20. Set the plug upside down on the counter.
21. Place the chisel next to the exposed vane and begin chipping downward
away from the vane.
22. Proceed around the bob by removing the cement from between the
vanes.
Procedure 2:
Tools Needed
- 10" Crescent Wrench
-
⅝" Wrench
- Channel Locks
- Soft Face Hammer
- Cell Cap Wrench
- Heavy Duty Vice
- Small Sledge Hammer
- Rock Chisel
- Knock Out Tool
1. Secure the SGSM cell in a heavy duty vice.
2. Use a crescent wrench to hold the bob shaft housing in place.
3. Unscrew the magnetic housing on top of the cell.
4. Remove the magnetic housing.
5.
Use a ⅝" wrench to loosen the thermocouple connection and pull it
straight up to make sure it pulls clear of the set cement.