When Multiple Conditioning function is checked
OFF
, “Strength #5”, will
prompt the software to turn on the alarm, and stop the test.
Multiple Conditioning Turned
OFF
Multiple Conditioning Turned
ON
Multiple Conditioning
- When this option is turned
ON
, the software will
measure gel strength periodically until the cement reaches the value in the
“Strength #5” field entered in the setup screen. Then it will condition the
cement for the time period specified in the “Conditioning Time” field on the
Main Setup screen. This cycle will continue until the gel strength reaches the
value in the “Final Strength” field on the Test Info Setup screen. Once the gel
strength reaches the “Final Strength, the alarm will sound and the test will end.
This value is used for all conditioning periods when “Multiple Conditioning” is
on. If no value is specified, the default is 30 minutes.