OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
11
Setup
1.
Carefully remove the instrument from the crate and place it on a flat,
stable surface. Make sure to allow for adequate air flow around the unit,
especially near the vents on the sides.
To ensure personal safety, always use proper lifting techniques.
Position the unit so that the user can quickly disconnect plugs in case of
an emergency.
2. Make sure all switches are off and the pressure relief knob is turned
completely counterclockwise.
3. Connect the “Air Supply”, “Water Supply”, “Drain”, “Coolant Supply”, and
“Coolant Drain” connectors on the back of the unit to their appropriate
source and plug in the power cords.
Make sure the “Coolant Drain” is suitable for high temperature discharge.
4. Connect the UCA/SGSM -to a PC with the supplied cables (USB for the
UCA and RS-232 for the SGSM).
5. Connect the SGSM head to the UCA with the supplied LPT cable.
6. Turn the “Main Power” switch on.
7. Put the “UCA/SGSM” switch to the correlating position of the desired test.
PC
S
G
S
M
UCA/SGSM
Cell
RS-232 (For SGSM)
LPT
USB
(For UCA)