Quick Start
SGSM
1. Set up the software with the desired parameters (see page 35).
2. Prepare the cell caps and cell body.
3. Assemble the bottom cell cap. See page 44.
4. Install the bottom cell cap to the bottom of the cell body. See page 46.
5. Assemble the top cell cap. See page 47.
6. Assemble the SGSM shaft assembly and install it to the top cell cap.
See page 49.
7. Install the top cell cap to the cell assembly.
8. Install the test cell to the cabinet.
9. Connect the SGSM head to the shaft assembly and run a drag check.
See page 52.
10.
Mix the slurry and fill the cell. See page 53.
11. Connect water supply line, thermocouple, and transducer to the test cell.
See page 54.
12. Run the test. See page 55.
13. Allow test cell to cool down. See page 56.
14. Evaluate the test data. See page 57.
15. Remove test cell.
16. Disassemble and clean the cement from the test cell. See page 58.
17. Clean and inspect all of the test cell components.