CURR?
This query returns the presently programmed current level of the power supply.
Return value
"current"
ex) curr?
return value "4.50"
CURR:STEP {<numeric value>}
This command sets the step size for current programming with the CURRent UP and
CURRent DOWN commands.
Program the power supply to the allowed setting value resolution.
Set the voltage to 0.5A
CURR:STEP?
This query returns the value of the step size currently specified. The returned parameter is a numeric value.
Return value
"numeric value"
ex) curr:step?
return value "0.50"
CURR:OCP {<numeric value>}
This command sets the current level at which the overcurrrent protection circuit will trip.
Input current within OVP setting range.
Set the OVP Level to 5.2A
CURR:OCP?
This query returns the overcurrent protection trip level presently programmed.
Return value
"numeric value"
ex) curr:ocp?
return value "5.20"
CURR:OCP:TRIP?
This query returns a ‘‘1’’ if the overcurrent protection circuit is tripped and not cleared or
a ‘‘0’’ if not tripped.
Return value
"0" - not tripped state.
"1" - OCP Trip state and cut the output.
ex) curr:ocp:trip?
return value "1"
CURR:OCP:CLE
This command causes the overcurrent protection circuit to be cleared.
Before release trip, refer to "3-4. Programming Over Current Protection(OVP)" in other to check
the reason of trip occuring and remove this reason.
Clear the OCP Trip
When appears OCP Trip, cut the voltage/current output. At that time, if you set the voltage/current,
it can be set. But DC output is still cut unless release the trip mode.
> numeric value
ex) curr:step 0.5
> numeric value
ex) curr:ocp 5.2
ex) curr:ocp:cle