3-13. Store
You can store up to ten different operating states(Output ON/OFF state, Voltage, Current, OVP,
OCP, UVL, OVL, UCL, OCL) in non-volatile memory.
The following steps show you howto store an operating state.
Store Key can be used to calibrate the power supply.
About calibration, refer to "4. CALIBRATION" .
▌
Turn on the power supply
After turned on, check the power supply displays"**OUTPUT OFF**" Mode.
▌
In order to store operating state in non-volatile, press the "Save" key
LCD Display STORE NO, 01
▌
Turn the encoder knob to specify the memory location among 01 ~ 10.
For examples, turn the knob to the right to specify the memory location 2
LCD Display
STORE NO, 02
▌
In order to store in your selected memory location 2, press the "Save" key.
LCD Display
Saved
▌
Appear "Saved" message, and return to previous state
≫ Related Remote Interface Command
*SAV {1|2|3|4|5|6|7|8|10}
apply: *SAV 4
Store operating states at location 4 in Non- volatile memory.
If you store operating state when the power supply output is on. it may give rise to serious trouble
when you recall, So. It is safe to store under the output off mode.
You can initialize the non-volatile memory. Refer to "5-3. USER -MEM CLEAR"
Note
- 50 -
Note