Current Calibration
▌
Send the "ON" command to electronic load.
▌
Send the power supply current minimum calibration command.
Sending command
▌
The mode of electronic load is CC mode, so set the current value is higher than the
power supply's maximum output ampere value. If you send the current calibration value
is lower than the power supply's maximum ampere value, this is CV mode and
wrong calibration value is measured and saved at the power supply. So remind to set
the sending value is higher than power supply's maximum output ampere.
▌
After waiteing for minimum 5 minutes more, measures DVM's voltage that is connected
with current monitoring resistor.
▌
Send calculated current value that caculated by current value calculating program prepared.
For examples, if calculated current value is 0.1234, sends as following.
Sending command
▌
Send current maximum calibration command of the power supply.
Sending command
▌
After waiting for minimum 5 minutes more, measures DVM's voltage that is connected
with current monitoring resistor.
▌
Send calculated current value that calculated current value calculating program prepared.
For examples, if calculated current value is 60.553, sends as following.
Sending command
CAL:CURR MAX
CAL:CURR 60.553
CAL:CURR 0.1234
CAL:CURR MIN
- 62 -
Содержание EX 100-12
Страница 1: ...Programmable DC Power Supply User Manual www odacore com EX Series Manual Part NO 019EX 03...
Страница 4: ...Safety Symbol...
Страница 91: ...92...
Страница 92: ...93...