3-15. Output ON/OFF
Enables or disables the power supply output. This key toggles between on and off.
Therefore you can acquire the effect of output on/off without remove the connected DUT.
Proceed as follows.
▌
Turn on the power supply
After turned on, check the power supply displays"**OUTPUT OFF**" Mode.
▌
Basic state is output off mode. In order to enable output on,
press the "OUTPUT ON/OFF" key.
▌
In order to enable output off, press the "OUTPUT ON/OFF" key once more.
≫ Related Remote Interface Command
OUTP {OFF|ON}
OUTP?
Apply: Check output state and in case output is off, the method of changing to output on.
Check the output state (Value"0" means output off)
OUTP ON
If you turn on the power supply and the default of output is not a output off state,
enter to "Factory" Mode and go to "Last State" mode in order to change to disable of "Last State"
And then turn off the power supply and turn on the power supply. You can see the appearing the
"OUTP OFF" display.
Note
- 52 -
OUTP? Return value "0"