When occurs OVP trip, "OVP TRIPPED" message appears on the display
▌
In order to clear tripped OVP, increase the OVP Level.
In order to increase OVP level, press the "Protection" key
LCD Display
protect> 1.ovp
▌
In the "protect> 1.ovp ", press the "Protection" key one more time.
LCD Display
ovp set> 15.00v
▌
Move the under bar cursor where you want to change the level.
▌
Readjust the OVP level by using the encoder knob.
▌
Finished to change the level, press the "Protection" key
LCD Display
CHANGED
After displayed above message, return to before state that tripped OVP.
≫ Related Remote Interface Command
VOLT:OVP {<voltage>}
VOLT:OVP?
VOLT:OVP:TRIP?
VOLT:OVP:CLE
apply 1: Method of checking tripped OVP or not and re-adjust the OVP Level.
VOLT:OVP:TRIP? Return value "1"
Check the OVP Trip state
VOLT:OVP 32
Set the OVP Level
VOLT:OVP:CLE
Clear OVP Trip
apply2: Method of checking tripped OVP or not and decrease the OVP Level.
VOLT:OVP:TRIP? Return value "1"
Check the OVP Trip state
VOLT 0
Set the voltage value to zero
VOLT:OVP:CLE
Clear OVP Trip
- 32 -