FACT:LOAD-DEF
This function is to initialize to the factory mode's setting value.
You can clear in User memory field excepted calibration data.
Initialize to the factory mode's setting value
6-8. System Commands
This is the commamd that is various control command of the power supply.
SYST:BEEP
The event to beep sound of power supply
ex) syst:beep
Make beep sound.
SYST:ERR?
Confirm the power supply's error that occurred.
Error recording, save in volatile memory up to 10 event. If the error occurs more than 10, erase
the first occurred error at first. If check the error, automatically erase the comfirmed error and there
are no error, you can r0 value,”No error” Message.
Return value - error number ,"message"
ex) syst:err?
return value -222,"Out of data"
1. About errer, refer to"7. Error Messages".
2. If you send "CLS" command, every error is cleared.But in case of "*RST" command, not cleare.
SYST:VERS?
You can confirm the power supply's version.
Return val
"YYYY.Ver"
YYYY - developed year.
Ver - Product version of development year.
ex) syst:vers?
return value "2008.3"
OUTP {OFF|ON}
You can control the power supply's DC output state of ON or OFF.
> ON
Output on.
> OFF
Output off.
ex1) outp on
DC power output on
ex2) outp off
DC power output off
ex)fact:load-def
- 84 -