Enters a voltage value that you obtained by reading an external DVM meter
If the value is 31.607V, enters as following below.
LCD Display
V-high 31.61v
This is an examples for 30V
▌
After finishing to enter the value, press the "Calibration" key.
LCD Display
ADC datA f100h
About 500ms, display HEX value instead
of readback value.
LCD Display
2.cal-volt high
Return to the previous state of calibration
▌
Volatage caliration is finished. It can be applied next operation of power supply.
Current Calibration
▌
Turn off all device and connect all device before calibration
>When current calibration, connect the power supply (+) output terminal and electronic load
and connect the power supply (-) output terminal and shunt lead for current monitoring
and then connect shunt's the other lead to electronic load (-) terminal
>Connect the DVM (+) test lead at the electronic load connction of current monitoring resistor
connect the DVM (-) test lead at the other side of current monitoring resistor
▌
Turn on Power switch in a pessing calibration key.
▌
Power switch on.
▌
After self test is finished, and when appearing "1.cal-volt low" message,
release calibration key.
▌
Move to "3.cal-curr low" mode by using cursor key and press the "Calibration" key
LCD Display
a-LOW 00.14a
Display current can be different
Wait for 5 minute
▌
Wait until DVM display voltage becomes stable.