NI 4071 7½-Digit FlexDMM Calibration Procedure
d.
Call the niDMM Config Measurement VI again, changing
Function
to
NIDMM_VAL_AC_VOLTS_DCCOUPLED
, and select the
Resolution in Digits
instance.
e.
Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
: Set as shown in
Table 8 for the current iteration
Measurement_Function
:
NIDMM_VAL_AC_VOLTS_DCCOUPL
ED
Range
: Set as shown in Table 8 for
the current iteration
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 21.
Maximum_Time
:
–1