NI 4071 7½-Digit FlexDMM Calibration Procedure
27. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
28. Call the niDMM Read VI.
29. Generate 200 V at 30 Hz on the calibrator.
30. Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_AC_VOLTS_DCCOUPL
ED
Range
:
700
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. The DMM must be in
the 700 V range before you apply the
voltage.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 21.
Maximum_Time
:
–1