NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
39
14. Repeat step 13 for each of the remaining iterations shown in Table 4.
15. Generate 450 mV at 30 Hz on the calibrator.
Table 4.
niDMM Config Measurement Parameters
Iteration
Calibrator Output
Signal Parameters
niDMM Config Measurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
45 mV
50 Hz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
50 Hz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5
2
45 mV
1 kHz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5
3
45 mV
1 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
45 mV
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
4
45 mV
20 kHz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
20 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5
5
45 mV
50 kHz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
50 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5
6
45 mV
100 kHz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
100 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5
7
45 mV
300 kHz
NIDMM_VAL_AC_
VOLTS
0.05
6.5
45 mV
300 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.05
6.5