NI 4071 7½-Digit FlexDMM Calibration Procedure
47. Refer to Table 16 for the appropriate calibrator output signal and parameter measurement
values as you complete the following steps:
a.
On the calibrator, generate the value listed under Calibrator Output in Table 16 for the
current iteration.
b.
Call the niDMM Read VI.
48. Repeat step 47 for each of the remaining iterations shown in Table 16.
49. Generate 2.2 A on the calibrator with the current output set to
NORM
.
50. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 25.
Maximum_Time
:
–1
Table 16.
niDMM 1 A Linearity Settings
Iteration
Calibrator Output
1
0.5 A
2
0.25 A
3
–
0.25 A
4
–
0.5 A
5
–
0.9 A