![National Instruments FlexDMM PXI-4071 Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/national-instruments/flexdmm-pxi-4071/flexdmm-pxi-4071_calibration-procedure_709945061.webp)
NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
61
Verifying 2-Wire Resistance
To verify the 2-wire resistance of the NI 4071, complete the following steps:
1.
Plug in the insulated banana plug shorting bar across the
HI
and
LO
banana plug connectors
on the NI 4071.
2.
Wait one minute for the thermal EMF to stabilize.
3.
Call the niDMM Reset VI.
4.
Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_close
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the
following parameter:
Instrument_Handle
:
The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7
.
5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e6