14
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
19. Use a writable niDMM property node to set the input resistance of the NI 4071 to >10 G
20. Call the niDMM Read VI.
21. Use a writable niDMM property node to set the input resistance of the NI 4071 to 10 M
.
LabVIEW Block Diagram
C/C++ Function
Call
Call
niDMM
SetAttributeVi
Real64
with the
following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPU
T_RESISTANCE
Attribute_Value
:
NIDMM_VAL_
GREATER_THAN_
10_GIGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1