NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
43
19. Generate 4.5 V at 30 Hz on the calibrator.
Table 5.
niDMM Config Measurement Parameters
Iteration
Calibrator Output
Signal Parameters
niDMM Config Measurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
450 mV
50 Hz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
50 Hz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
2
450 mV
1 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
3
450 mV
1 kHz
NIDMM_VAL_AC_
VOLTS
5
6.5
450 mV
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
5
6.5
4
450 mV
20 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
20 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
5
450 mV
50 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
50 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
6
450 mV
100 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
100 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5
7
450 mV
300 kHz
NIDMM_VAL_AC_
VOLTS
0.5
6.5
450 mV
300 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
0.5
6.5