NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
51
Table 7.
niDMM Config Measurement Parameters
Iteration
Calibrator Output
Signal Parameters
niDMM Config Measurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
45 V
50 Hz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
50 Hz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5
2
45 V
1 kHz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5
3
45 V
1 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
45 V
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
4
45 V
20 kHz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
20 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5
5
45 V
50 kHz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
50 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5
6
45 V
100 kHz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
100 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5
7
45 V
300 kHz
NIDMM_VAL_AC_
VOLTS
50
6.5
45 V
300 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
50
6.5