NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
115
59. Call the niDMM Cal Adjust Offset VI with the following parameters:
60. Remove the Fluke 5440 cables from the NI 4071.
61. Plug in the insulated shorting bar across the
HI
and
LO
banana plug connectors of the
NI 4071.
62. Wait two minutes for the thermal EMF to stabilize.
63. Call the niDMM Cal Adjust Offset VI with the following parameters:
64. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
100
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
1e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA