82
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
16. Connect the Fluke 5720A calibrator to the NI 4071 with a DC current shunt/voltmeter
system in series, as shown in Figure 5. Table 14 lists the cable connections. Use the shunt
recommended for ±100 µA as indicated in Table 13.
17. Wait two minutes for the thermal EMF to stabilize.
Figure 5.
Shunt/Voltmeter Connections for Current
18. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
1
NI 4071
2
Fluke 5720A Calibrator
3
Precision Shunt Resistor
4
Voltmeter
Table 14.
Shunt/Voltmeter Connections
Banana Plug
Connector
(NI 4071)
Precision
Shunt
Resistor
Voltmeter
Binding Post
(Fluke 5720A Calibrator)
—
HI
HI
OUTPUT HI
HI SENSE
LO
LO
—
LO
—
—
OUTPUT LO
—
GND
V GUARD
V GUARD
HI
LO
HI
LO
HI
HI
LO
HI
LO
HI
AUX
CURRENT
GUARD
GROUND
2
HI
LO
AUX
I/O
300V
MAX
5V
MAX
AMPS
NI PXI-4071
7
1
/
2
-Digit FlexDMM
500V MAX
CAT I
HI
LO
1kV
MAX
3A, 250V
MAX
INPUT
1kV
MAX
1
V
4
3