NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
75
5 G
Range Verification
This verification procedure is optional and requires additional test equipment. If you do
not
want
to verify the 5 G
range, you have completed verifying the 2-wire resistance of the NI 4071.
1.
Connect a 1 G
standard resistor to the NI 4071 using the Fluke 5440 cable or applicable
low thermal cabling, as shown in Figure 4. Table 12 lists the cable connections.
Figure 4.
Cable Connections for 1 G
Standard Resistor
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_close
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Table 12.
Fluke 5440 Cable Connections
Banana Plug
Connector
(NI 4071)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post Label
(1 G
Standard Resistor)
HI
Red
HI
LO
Black
LO
(No connection)
Blue
GND
HI
LO
AUX
I/O
300V
MAX
5V
MAX
AMPS