18
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
32. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
33. Use a writable niDMM property node to set the input resistance of the NI 4071 to >10 G
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
0.1
LabVIEW Block Diagram
C/C++ Function
Call
Call
niDMM
SetAttributeVi
Real64
with the
following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPU
T_RESISTANCE
Attribute_Value
:
NIDMM_VAL_GREAT
ER_THAN_10_GIGA
OHM