NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
33
70. Call the niDMM Read VI.
71. Generate 1000 V on the calibrator.
72. Call the niDMM Read VI.
73. Generate –1000 V on the calibrator.
74. Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. The DMM must be in
the 1000 V range before you apply the
voltage.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1