NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
111
43. Call the niDMM Cal Adjust Gain VI with the following parameters:
44. Generate 100 k
on the calibrator with external sense turned on and 2-wire compensation
turned off.
45. Call the niDMM Cal Adjust Gain VI with the following parameters:
46. Generate 0
on the calibrator with external sense turned on and 2-wire compensation
turned off.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
The value that is
displayed on the calibrator for
1.9 M
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
100e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
The value that is
displayed on the calibrator for 100 k