NI 4071 7½-Digit FlexDMM Calibration Procedure
7.
Call the niDMM Cal Adjust Gain VI with the following parameters:
8.
Generate –1
A on the calibrator with the current output set to
NORM
.
9.
Call the niDMM Cal Adjust Gain VI with the following parameters:
10. Generate 10
A on the calibrator with the current output set to
NORM
.
11. Call the niDMM Cal Adjust Gain VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1e–6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
1e–6
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1e–6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
–1e–6
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
10e–6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
10e–6