NI 4071 7½-Digit FlexDMM Calibration Procedure
35. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
36. Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7
.
5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e3
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that the
reading or measurement value falls
between the tolerances listed in
Table 24.
Maximum_Time
:
–1