NI 4071 7½-Digit FlexDMM Calibration Procedure
10. Call the niDMM Read VI.
11. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
12. Call the niDMM Read VI.
13. Generate 0.9 mA at 1 kHz on the calibrator with the current output set
to
NORM
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 26.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
6
.
5
Measurement_Function
:
NIDMM_VAL_AC_CURRENT
Range
:
1e–3
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 26.
Maximum_Time
:
–1