NI 4071 7½-Digit FlexDMM Calibration Procedure
You have completed verifying the DC current of the NI 4071. Select one of the following
options:
•
If you want to continue verifying other modes, go to the
•
If you do
not
want to verify other modes
and
you are performing a
post-adjustment
verification, go to the
Completing the Adjustment Procedures
section.
•
If you do
not
want to verify any additional modes
and
you are performing a
pre-adjustment
verification, call the niDMM Close VI to close the session.
Verifying AC Current
To verify the AC current of the NI 4071, complete the following steps:
1.
Reset the calibrator.
2.
Connect the NI 4071 and the Fluke 5720A calibrator using the Fluke 5440 cable, as shown
in Figure 6. Table 15 lists the cable connections.
3.
Call the niDMM Reset VI to reset the NI 4071 to a known state.
4.
Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_close
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the
following parameter:
Instrument_Handle
:
The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
6
.
5
Measurement_Function
:
NIDMM_VAL_AC_CURRENT
Range
:
100e–6