NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
13
17. Call the niDMM Read VI.
18. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeViRea
l64
with the following
parameters:
Instrument_Handle
:
The instrument handle
from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT_
RESISTANCE
Attribute_Value
:
NIDMM_VAL_10_MEGA
OHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function.
Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
10