76
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
2.
Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
3.
Call the niDMM Control VI.
4.
Wait two minutes for the thermal EMF to stabilize.
5.
Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7.5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
:
5e9
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the
following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Control Action
:
Commit
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the
tolerances listed in Table 24.
Maximum_Time
:
-1