NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
117
68. Call the niDMM Cal Adjust Offset VI with the following parameters:
69. Call the niDMM Cal Adjust Miscellaneous VI with the following parameters:
You have completed adjusting the DC voltage and resistance modes of the NI 4071. Select one
of the following options:
•
If you are performing additional adjustments, refer to the following sections, as applicable:
–
Adjusting AC Voltage (AC- and DC-Coupled) Modes
–
•
If you are
not
performing additional adjustments, refer to one of the following sections:
–
—to verify your new calibration coefficients before saving
them to the EEPROM
–
Completing the Adjustment Procedures
—if you do
not
want to verify the adjustments
you have just made
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustMisc
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Type
:
NIDMM_EXTCAL_MISCCAL_SECTI
ON