NI 4071 7½-Digit FlexDMM Calibration Procedure
36. Call the niDMM Read VI.
37. Generate –90 mV on the calibrator with the range locked to 2.2 V.
Note
This calibrator range prevents the 50
output resistance of the 220 mV
range from creating a voltage divider with the internal resistance of the NI 4071.
38. Use a writable niDMM property node to set the input resistance of the NI 4071 to >10 G
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Subtract the
previously stored 100 mV 10 M
mode offset value from this
measurement. Verify the reading or
measurement value falls between the
limits listed in Table 20.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function
Call
Call
niDMM
SetAttributeVi
Real64
with the
following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPU
T_RESISTANCE
Attribute_Value
:
NIDMM_VAL_GREAT
ER_THAN_10_GIGA
OHM