![National Instruments FlexDMM PXI-4071 Скачать руководство пользователя страница 55](http://html.mh-extra.com/html/national-instruments/flexdmm-pxi-4071/flexdmm-pxi-4071_calibration-procedure_709945055.webp)
NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
55
Table 8.
niDMM Config Measurement Parameters
Iteration
Calibrator Output
Signal Parameters
niDMM Config Measurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
630 V
50 Hz
NIDMM_VAL_AC_
VOLTS
700
6.5
630 V
50 Hz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
2
630 V
1 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
630 V
1 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
3
200 V
20 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
200 V
20 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
4
200 V
50 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
200 V
50 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
5
200 V
100 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
200 V
100 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5
6
70 V
300 kHz
NIDMM_VAL_AC_
VOLTS
700
6.5
70 V
300 kHz
NIDMM_VAL_AC_
VOLTS_DCCOUPLED
700
6.5