System
Installation
GMW-25
2-35
2.12 System
Check
Before the GMW-25 generator can be used, it is imperative that the
following procedure be followed to ensure optimal and safe operation.
1. If you've purchased your generator with the optional solenoid valve
kit (as mentioned in section 2.8.3), make sure the solenoid plug is
attached to the solenoid connector on the rear panel marked
“SOLENOID.”
2. Check that the water hoses are securely connected to the water
connectors on the rear panel. (See section 2.10 for more
information on the connectors.)
3. Be certain the external interlock plug is inserted in the interlock
connector on the rear panel marked “INTERLOCK.” AC Power will
not engage until this interlock is satisfied (plugged). (See section
2.7 for more information about the safety interlocks.)
4. Assure that the AC line cord is plugged into 208 VAC, 3-phase
power. (See section 2.9 for more information on AC Power
Requirements.)
5. Make sure the RS-232 serial cable is connected to the 9-pin,
D-type serial connector on the rear panel.
6. Check that the Analog Interface is connected to the 25-pin, D-type
analog I/O connector on the rear panel. (See section 2.4 for more
information on the types of analog interfaces available.) This may
vary depending upon the type of analog interface that was ordered.
7. Be sure the RF output cable is connected to the RF Output
connector on the rear panel.
8. Make certain the RF output cover is on, defeating the RF interlock.
AC Power will not engage until this interlock is satisfied (plugged).