System Installation
2-30
GMW-25
Rear Panel Drawing
Figure 2.8.3 shows the rear panel for the GMW-25 Generator and calls
out each of the connectors as described in the table to the left. In order
to maintain EMC compliance, wires to external interlock plugs (Ext.
Interlock, 24 V Solenoid Valve, and Water Flow Readback) must be
wound 3 times through a suitable ferrite assembly [FerriShield Part No.
SS28B2033 (or equivalent)] positioned as close to the rear panel as
possible.
GMW-25 Rear Panel View
Figure 2.8.3