6.3.2
Switch to calibration mode
Stand-by condition
[CFUNC]
6.3.3
Set the scale interval
[CCAL]
[SCAL]
[
✶✶
] is the currently saved scale interval.
[TARE WEIGHT] [NET-GROSS WEIGHT]: Selects the scale interval
[SET]: Interrupts the setting and returns to the CCAL display
[F]: The displayed value is memorized. Proceed to the next step.
The scale interval to be used
[DISP]
▼ Set maximum load
1. Press the [ON-OFF] key to move from normal weighing mode to the stand-by
condition.
2. Press the [ON-OFF] key while pressing the [SET] key.
[CFUNC] is displayed.
3. Press the [NET-GROSS WEIGHT] key.
[CAL]
[CAL] is displayed.
4. Press the [F] key.
[CCAL]
▼ Set the scale interval
[CCAL] is displayed, and it enters calibration mode.
1. When [F] key is pressed twice from the [CCAL] display, [D-
✶✶
] is displayed.
2. Select the scale interval from 1, 2, 5, 10, 20 and 50.
3. Press the [F] key after setting
Weight indicator CSD-903
6 Calibration
EN-38
Minebea Intec