6.3.4
Set the maximum load
[DISP]
Saved maximum load
If the maximum load has already been changed, the currently saved maximum load is
displayed.
[TARE WEIGHT] [NET-GROSS WEIGHT]: The changed digit value adjusts.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[ZERO]: The value changes to [0].
[SET]: Interrupts the setting and returns to the CCAL display
[F]: The displayed value is saved. Proceed to the next step.
Changed maximum load
[LOAD]
▼ Set the weight mass
6.3.5
Setting the mass of the weight
[LOAD]
[
✶✶✶✶
] is the maximum load.
Weight mass equal to maximum load.
Set the mass of the weight on the load cell
The same setting as the maximum load is used if there is a maximum load value.
[TARE WEIGHT] [NET-GROSS WEIGHT]: The changed digit value adjusts.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[ZERO]: The value changes to [0].
[SET]: Interrupts the setting and returns to the CCAL display
[F]: The displayed value is saved. Proceed to the next step.
1. When [F] key is pressed from the [DISP] display, [01000] is displayed.
2. Press the [F] key after setting
1. When [F] key is pressed from the[LOAD] display, [
✶✶✶✶
] is displayed.
6 Calibration
Weight indicator CSD-903
Minebea Intec
EN-39
Содержание CSD-903
Страница 243: ......