18.2 BCD output
18.2.1
Speciications (model CSD-903-P15)
The BCD output speciications are given below.
Speciications
I/O
Content
BCD data
Output
8 digit parallel output
POL. (Polarity)
Output
ON for negative polarity output, OFF for positive po-
larity output
P.C. (Print com-
mand)
Output
ON for a certain length of time after BCD output is
converted
ERROR
Output
ON when the error occurs.
OVER
Output
ON synchronizing with the OL display (Over load).
Decimal point 10
1
,
decimal point 10
2
,
decimal point 10
3
,
decimal point 10
4
Output
ON synchronizing with the decimal point display
setting.
Stable
Output
ON when stability is detected.
Gross value
Output
ON when the BCD output is gross weight
HOLD
Input
BCD output hold
BCD-ENABLE
Input
BCD-related output is forcibly set to high impedan-
ce.
18.2.2
Operation mode of BCD output
The operation mode of BCD output is selected by the setting of function F-30.
The operation mode of BCD output can be selected from [Stream mode], [Synchronizing
with print], [Synchronizing with accumulation] and [Synchronizing with inish].
Default is set to [Stream mode].
18.2.3
BCD output target
The BCD output target is selected by the function F-31 setting.
Function No.
Set value
Contents
0
Display interlock
1
Gross weight
2
Net weight
3
Tare weight
4
Accumulated total value
F-31
5
Accumulated count
The default is [0].
18 Options
Weight indicator CSD-903
Minebea Intec
EN-169