Item
Function No.
Number Return command set value
Discharge monitor timer
72
46
0000 to 9999 (Unit: 0.01 s)
Discharge gate close delay time 72
47
0000 to 9999 (Unit: 0.01 s)
Discharge inish output time
72
48
0000 to 9999 (Unit: 0.01 s)
Tare weight cancellation at
discharge inish
72
49
0: Invalid; 1: Efective
Note:
Change [7*] of abovementioned command to [8*] at the time of writing.
Items [40] to [47] are only read-only.
When calibration lock is valid, command [71] is read-only.
The data setting values [90] to [97] are read-only as they are calibration items.
Writing switches to commands [7*] to [8*].
18.3.12.13 Communication error processing
The device replies with the error command to the host side when a communication or
execution error occurs.
6
6
8
4
2
@
0
0
E
R
0
1
Terminator
Address
00 ~ 31
Error command No.
Error
Command No.
Details
Remarks
01
Non-executable error
In calibration mode, function mode or check mo-
de
02
Other errors for this unit
The receiving command cannot be executed.
10
Parity error
Parity detection error
11
Fleming error
Stop bit detection error
12
Overrun error
Receiving command read error
13
Data code and data length error
The receiving data code and data length do not
correspond.
14
No pertinent command
The receiving command does not correspond.
20
Error display for zero calibration
The load cell output voltage is –2.5 mV/V or less.
21
Error display for zero calibration
The load cell output voltage is 2.5 mV/V or more.
22
Error display for span calibration
The diference in load cell output voltage bet-
ween span point and zero point is 0.0 mV/V or
less.
Weight indicator CSD-903
18 Options
EN-202
Minebea Intec