Item
Function No.
Number Return command set value
Lock key 1
70
08
0000 to 1111 (0 : Invalid; 1: Valid)
10
0
: ON/OFF key
10
1
: Setting key
10
2
: Accumulation key
10
3
: Tare weight set key
Lock key 2
70
09
0000 to 1111 (0 : Invalid; 1: Valid)
10
0
: Tare weight cancellation key
10
1
: Net or gross weight key
10
2
: Zero set key
10
3
: F key
Stability detection data width
70
10
000 to 999 (Unit: 0.1 D)
Stability detection time
70
11
00 to 99; Unit: 0.1 s)
Preset tare weight cancellation
operating condition
70
15
0: Efective; 1: Invalid
Automatic print operation
70
16
00 to 11
10
0
: Manual print operation only
when stable
0: Efective only when stable
1: Always efective
10
1
: Automatic print operation
0: Automatic printing of
1: Automatic printing only when stable
Hold operation
70
17
0: Synchronized with Hold
1: Synchronized with Inlow Finish
Hold target
70
18
0000 to 1111 (0: Invalid; 1: Valid)
10
0
: Display
10
1
: Comparison result
10
2
: CC-Link interface
10
3
: Serial interface
18 Options
Weight indicator CSD-903
Minebea Intec
EN-193