Note:
Data is output at constant intervals in stream mode. Host-side control is disabled.
Command mode
Data is sent back from CSD-903 to the host according to the command/data sent to
CSD-903 by the host (personal computer, sequencer, etc.). Execute communication
according to the following procedure.
START
Command/data
Data
Host→CSD-903
CSD-903→Host
Stream mode
When RS-232C or RS-422/485 are used, this mode continues to output the latest data of
the output target selected by function F-41. However, the output frequency depends on
the baud rate setting.
Synchronized with print
When RS-232C is used, output target data selected by function F-41 is output
synchronized with the print signal (F key, external control input and automatic print).
Note:
-
Communication can only be executed in weighing mode. An error code is
transmitted if other modes are used.
-
This mode does not respond in stand-by condition.
-
CSD-903 has no low control.
-
CTS/RTS signal is not used.
-
X-low control is not executed by the device.
-
Communication operation is an interactive mode.
18.3.4
Output target synchronized with RS-232C print for RS-232C and RS-422/485
RS-232C and RS-422/485 output targets are selected by function F-41.
RS-232C and RS-422/485 output targets can be selected from [Display interlock], [Gross
weight], [Net weight] and [Conditional load display data].
The default is [Display interlock].
18 Options
Weight indicator CSD-903
Minebea Intec
EN-175