Standard RS-485 interfa-
ce
Baud rate: Can be selected from 1 200, 2 400, 4 800, 9 600,
19 200 and 38 400 bps
Data bit length: Can be selected from 7 bit and 8 bit
Parity bit: Can be selected from none, even and odd
Stop bit: Can be selected from 1 bit and 2 bit
Terminator: Can be selected from CR+LF and CR
Communication method: Half duplex
Synchronization method: Start-stop synchronous method
Address: Select one from 0 to 31
Communication data: ASCII code
Cable length: Approx. 1 km
Connectable units: 32 units maximum
Termination resistance: External
Data transfer mode: Command, Modbus, Stream, Synchroni-
zed with inish, Synchronized with accumulation, Synchroni-
zed with print
Internal circuit and common are connected.
BCD output
(Sold separately)
Output
8-digit BCD, parallel output with polarity (POL.) applied (ON
for negative output, and OFF for positive output); P.C. (Print
command) ON a ixed time after BCD output conversion is
complete
ERROR ON when various errors occur OVER, STABLE, Gross
weight: ON when BCD output is gross weight
Above are open collector output. VCE = 30 V DC, IC = 20 mA
DC at maximum input
HOLD: Holding BCD output; BCD-ENABLE: Compulsorily of
for BCD output (High-impedance)
Above are for level input and efective during bridge input of
100 ms or more.
RS-232C
(Sold separately)
Baud rate: Can be selected from 1 200, 2 400, 4 800, 9 600,
19 200 and 38 400 bps
Data bit length: Can be selected from 7 bit and 8 bit
Parity bit: Can be selected from none, even and odd
Stop bit: Can be selected from 1 bit and 2 bit
Terminator: Can be selected from CR+LF and CR
Communication method: Half duplex
Synchronization method: Start-stop synchronous method
Communication data: ASCII code
Weight indicator CSD-903
21 Appendix
EN-236
Minebea Intec
Содержание CSD-903