Check on options
↓
The analog output operation is
wrong.
→ Yes →
To the analog output check.
↓
No
↓
The BCD output is wrong.
→ Yes →
To the BCD output check.
↓
No
↓
RS-232C is wrong.
→ Yes →
To the RS-232C check.
↓
No
↓
RS-422 is wrong.
→ Yes →
To the RS-422 check.
↓
No
↓
RS-485 is wrong.
→ Yes →
To the RS-485 check.
↓
No
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.
Analog output check
↓
Correct connecting?
→ No →
↓
Yes
↓
Connect wires surely, refer to
Chapter
Voltage output load resistance is 5
K
Ω
or more. Current output load
resistance is 510
Ω
or less.
→ No →
Make the load resistance suitable
for the speciications.
↓
Yes
↓
Conirm the setting concerning the
analog output.
↓
Correctly set.
→ No →
↓
Yes
↓
Set correctly, refer to Chapter
Inform Minebea Intec about the
contens of failure and situation at
site in details.
Weight indicator CSD-903
20 Troubleshooting
EN-226
Minebea Intec