Under normal operation
Return (device → host)
@
2
4
6
8
10
12
14
16
0
0
9
0
0
9
Terminator
Set value
Number
Sign
Command No.
Address
00 ~ 31
Note:
Setting the ifth decimal point digit
Example: Input data =100001 → 1.00001 mV/V
Under abnormal operation
Return (device → host)
6
6
8
4
2
@
0
0
E
R
2
0
Terminator
Address
00 ~ 31
6
6
8
4
2
@
0
0
E
R
2
1
Terminator
Address
00 ~ 31
Note:
-
A command to register mV/V of ive digits after the decimal point for ZERO (No. 09)
and SPAN (No. 12) to boost calibration has been available since ROM Ver. 2.300.
When registering mV/V values for ZERO and SPAN, use the command that registers
mV/V ive digits after the decimal point for ZERO (No. 09) and SPAN (No.12).
-
Continue to use the current command if the instrument is incorporated in a system
that already uses mV/V four digits after the decimal point for ZERO (No. 08) and
SPAN (No. 11).
Note that error in the set digit number may occur if the command is changed to ive
digits after the decimal point to register ZERO point (No.9) and SPAN point (No.12)
as mV/V.