RS-422C check
↓
Correct connecting?
→ No →
↓
Yes
↓
Connect wires surely with external
equipment, refer to Chap-
ter
.
The address number is correctly
set.
→ No →
↓
Yes
↓
Set the address number correctly,
refer to Chapter
.
The communication protocol is
correctly set.
→ No →
↓
Yes
↓
Match the setting of function cor-
rectly to the speciication of the
host according.
Command/data is correctly set.
→ No. →
↓
Yes
↓
Use the correct command/data, re-
fer to Chapter
.
The host speciies the low control.
→ Yes →
↓
No
↓
Set the low control from [YES] to
[NONE].
The change of RS-422/485 is pro-
perly set.
→ No →
Make the change of RS-422/485
correctly.
↓
Yes
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.